SCPI Sequence for Continuous Measurements

Tutorial Steps

// reset the instrument settings
*CLS;*RST
*ESE 0;*SRE 0

// Meas function
:CONF:FREQ:BTB

// Input settings. Should be after CONF, because CONF sets up many subsystems
// including INPUT
:INP:LEV:AUTO OFF;:INP:LEV 0

// Pacing. Note: for FREQ:BTB measurement time is used as pacing. For
// PERIOD:BTB specify :TRIG:SOUR TIM;:TRIG:TIM 0.01
:SENS:ACQ:APER 0.01

// Packed format, Time Information On, use Intel Byte Order (LSB first)
:FORM PACK;:FORM:TINF ON;:FORM:BORD SWAP

// Turn off the display
:DISP:ENAB OFF

// Measure continuously
:TRIG:COUN 1;:ARM:COUN INF

// Set up SRQ on Message Available (MAV). This will generate SRQ when
// message is available in output buffer
*ESE 1;*SRE 16

// Clear SRQ-related registers by reading them
*SRE?

// read the response (library dependent call)
…// read STB (library dependent call)

// enable SRQ (library dependent call)

// initiate the measurement
:INIT

// read results continuously
loop
{
:FETC:ARR? MAX
// wait for SRQ (library dependent call)

// read STB to clear MAV SRQ (library dependent call).
// Please, note: STB should be read by system call, not by sending *STB?, as it will generate nested SRQ

// read and process response

}

// disable SRQ (library dependent call)


// reset the counter settings
*ESE 0;*SRE 0
*CLS;*RST

Please note: this sample can be implemented without SRQ (which is not recommended though), then SRQ initialization part can be omitted and no waiting-for-SRQ and reading STB is required in fetching loop (in this case make sure that appropriate bus timeout is set, otherwise you may end up with read timeout).