Spectracom - Essential Ingenuity

CNT-9x Firmware Release Notes 1.24

Firmware Version

1.24

Firmware Date

05/04/2009

Enhancements

  1. Fix for *RST+No Signal (TimeView connect/disconnect): when sending *RST during no signal and then waiting some time (up to 43 s) with no signal either – bus freezes and many other things (like switching between measuring functions) do not work.
  2. Fix for GET + Agilent GPIB card (GET + SRQ on Message Available in fastest mode sometimes failed with Agilent card).
  3. Fix for timeouts when measuring burst and no signal is applied: in such situation no timeout happens
  4. Change for HP Compatible mode: now in this mode FW accepts newline as program message terminator (doesn’t require EOI).

Was this information helpful?